Representative Standard Curve
• Reportable range: 10 - 3160 pg/mL. This is the concentration range in which measurement of the analyte can be done with the highest precision, accuracy, and linearity.
• Sensitivity: The limit of detection of this assay is 6 pg/mL. This is the analyte concentration with absorbance two standard deviations higher than the zero standard.
• Accuracy: No international standard exists for calibration.
• Recovery: A mid-curve recovery of 85 - 121% was determined by spiking defined amounts of analyte standard into serum or plasma samples in repeated experiments.
• Precision: The intra-assay precision of this assay is 3.7% (CV). The inter-assay precision of this assay is 6.8% (CV).
Find supporting information and directions for use in the Product Information Sheet or explore additional protocols below.